VIBROTECH 300 is a robust 305 mm vibration-polishing platform intended for very gentle metallographic super-finishing. Its low-aggression action is suited to specimens that must retain inclusions and avoid preparation-induced surface hardening before EBSD or AFM examination.

ManufacturerPRESI
ModelVIBROTECH 300
Reference68100
Machine typeVibrating polishing machine
Polishing diameter305 mm
Finishing methodUltra-soft vibration polishing
Primary analytical usesEBSD; AFM
Real-time controlsVibration frequency and amplitude
Preparation effectsPreserves inclusions; removes surface hardening

Process role

Vibration polishing uses controlled motion of the polishing surface to finish specimens with low mechanical aggression. The method is positioned after conventional preparation when the final surface must be suitable for sensitive microstructural analysis.

The 305 mm working diameter supports a laboratory-scale batch or fixture arrangement appropriate to the selected preparation method. Frequency and vibration amplitude can be adjusted while the process is running, allowing the operator to tune motion to the specimen and polishing medium.

Applications and expected result

EBSD preparationFinal surface conditioning for electron backscatter diffraction, where deformation from earlier grinding and polishing can interfere with pattern quality.
AFM preparationGentle finishing before atomic-force microscopy, where surface relief and preparation damage can distort topographic measurement.
Inclusion preservationThe ultra-soft finishing action is intended to retain inclusions rather than pull them out during aggressive polishing.
Surface-hardening removalUsed as a final step to reduce the mechanically affected layer left by previous preparation stages.

Method development

Select the polishing medium, suspension and duration for the specimen material and analytical objective. Start from a conservative vibration setting, observe specimen movement and surface condition, then change frequency or amplitude one variable at a time.

Clean specimens and holders before the vibration stage so coarse abrasive from earlier steps cannot contaminate the final surface. Record settings, medium and time for repeatable laboratory work.

Care and safety

Confirm that the working surface and specimen fixtures are correctly seated before starting. Keep hands and loose items clear of the moving assembly and stop the machine before repositioning specimens.

Use polishing liquids only with appropriate chemical handling controls and personal protective equipment. Disconnect and isolate the machine before cleaning below the working surface or carrying out maintenance.