{
  "record_id": "0ca79687fcaf1ac7",
  "title": "QATM Qetch Electrolytic Polisher and Etcher Family Guide",
  "kind": "reference",
  "company": "QATM",
  "parameters": {
    "Manufacturer": "QATM",
    "Equipment family": "Qetch",
    "Models compared": "Qetch 100 M and Qetch 1000",
    "Shared process": "Electrolytic polishing and etching",
    "Shared maximum voltage": "90 V DC",
    "Qetch 100 M maximum current": "2 A",
    "Qetch 1000 maximum current": "14 A",
    "Qetch 100 M polishing area": "Approx. Ø9 mm",
    "Qetch 1000 program capacity": "Up to 200",
    "Qetch 100 M power source": "Li-ion storage battery",
    "Qetch 1000 connection power": "2 kVA"
  },
  "summary": "The QATM Qetch family covers two distinct metallographic preparation workflows: the battery-powered Qetch 100 M for localized field work and the higher-current Qetch 1000 for programmable laboratory polishing and etching. This guide separates the models by process scale, electrical envelope, control approach and installation needs.",
  "sections": [
    {
      "heading": "Choosing by work location",
      "paragraphs": [
        "Qetch 100 M is the mobile option. Its Li-ion battery, integrated process handpiece and approximately 9 mm polishing area support localized preparation on components outside the laboratory.",
        "Qetch 1000 is the laboratory option. The polishing and etching section is separated from the touchscreen controller so the process unit can be used in a fume cabinet."
      ]
    },
    {
      "heading": "Electrical and process scale",
      "rows": [
        [
          "Qetch 100 M",
          "160 W maximum power; 90 V DC; 2 A maximum current; approximately Ø9 mm polishing area"
        ],
        [
          "Qetch 1000",
          "2 kVA connection power; 90 V DC; 14 A maximum current; polishing 1 sec–25 min; etching 1 sec–5 min"
        ]
      ],
      "paragraphs": [
        "Both systems share a stated maximum voltage of 90 V DC, but their current ratings and intended work envelopes differ substantially. Parameters must be selected for the material, electrolyte and exposed sample area rather than transferred directly between models."
      ]
    },
    {
      "heading": "Control and method management",
      "paragraphs": [
        "Qetch 100 M uses a microprocessor-controlled automatic run, adjustable voltage, current and time, an LCD panel and a multicolor status LED on the handpiece.",
        "Qetch 1000 adds touchscreen control, a real-time current-voltage diagram and storage for up to 200 password-protectable programs. Its scan function supports method selection from the material's voltage response."
      ]
    },
    {
      "heading": "Physical planning",
      "rows": [
        [
          "Qetch 100 M",
          "380 × 355 × 180 mm; approximately 8.5 kg without electrolyte; 1 L electrolyte tank"
        ],
        [
          "Qetch 1000",
          "330 × 220 × 400 mm; approximately 7 kg; exchangeable electrolyte vessel"
        ]
      ],
      "paragraphs": [
        "The mobile model's stated weight excludes electrolyte, while the laboratory model's separated layout requires planning for both the process area and controller. Compare complete operating configurations rather than equipment-only dimensions."
      ]
    },
    {
      "heading": "Chemical and safety controls",
      "paragraphs": [
        "Electrolytic preparation requires a material-specific method, compatible electrolyte, secure electrical contact and controlled exposure time. Establish ventilation, personal protective equipment, spill containment, storage and disposal procedures before use.",
        "For Qetch 100 M field work, include transport containment and battery checks. For Qetch 1000 laboratory work, verify fume-cabinet placement and the cleaning procedure for the selected electrolyte system."
      ]
    }
  ],
  "updated_at": "2026-10-14 04:38:59"
}