QATM Qetch 1000 is a fully automatic electrolytic polishing and etching system for metallographic sample preparation. Its process unit is separated from the control unit, allowing the working section to be positioned in a laboratory fume cabinet while recipes and live electrical data are managed from the touchscreen.
| Manufacturer | QATM |
|---|---|
| Model | Qetch 1000 |
| Process | Electrolytic polishing and etching |
| Automation | Fully automatic |
| Control | Touchscreen with QATM software |
| Programs | Up to 200, password-protectable |
| Connection power | 2 kVA |
| Maximum voltage | 90 V DC |
| Maximum current | 14 A |
| Polishing duration | 1 sec–25 min |
| Etching duration | 1 sec–5 min |
| Dimensions | 330 × 220 × 400 mm |
| Weight | Approx. 7 kg |
| Process monitoring | Real-time current-voltage diagram |
| Cleaning | Water washing program |
Qetch 1000 removes material electrochemically for metallographic polishing and etching. The separated polishing/etching unit and controller support work inside a laboratory fume cabinet without placing the operator interface in the process area.
A material scan displays the current-voltage curve so the operator can identify a suitable process point before saving a method. The stated polishing window is 1 second to 25 minutes; the etching window is 1 second to 5 minutes.
The touchscreen control stores up to 200 parameter sets, with password protection available for saved programs. Real-time current and voltage display supports monitoring during the cycle.
Output voltage and process duration are variable. This makes the unit suitable for laboratories that need repeatable recipes across different materials while retaining the ability to tune a new process from its voltage response.
The current rating is substantially higher than the portable Qetch 100 M and should be considered together with the selected electrolyte, exposed sample area and material-specific method.
The electrolyte vessel is exchangeable and can be closed with a lid for storage. A water washing program is provided for unit cleaning.
Before operation, verify electrolyte compatibility, fume extraction, electrical limits, sample contact and site procedures for chemical handling. Use the required protective equipment and follow the laboratory's electrolyte storage, spill response and waste-disposal rules.