DLyte Titanium Acetabular Implant Polishing Case

A titanium acetabular implant case focused on polishing only the internal surface, reducing roughness below 0.05 µm with low material removal and controlled batch processing.

ComponentTitanium acetabular implant
Treated areaInternal surface only
ElectrolyteDry Suspension; L-130 for titanium
Process time45 minutes
Batch capacity2 pieces
Final roughnessBelow 0.05 µm
Reported material removal0.03 g per piece
Media bowl volume210 litres
Titanium removal capacity per media charge2 kg
Calculated treatment quantityMore than 65,000 implants per media charge

Finishing objective

The process was defined to polish the internal bearing surface while leaving the exterior treatment boundary controlled.

The documented target was roughness below 50 nm with homogeneous results between batches.

Validated cycle

Trial cycles varied time and other process parameters, with profilometer measurements taken in multiple areas. The selected cycle used dry-suspension media.

The recorded production condition processed two implants in 45 minutes and achieved final roughness below 0.05 µm.

Material-removal accounting

Production controls

POLISHX technical profile · 2026-10-13 13:31:12