  "record_id": "54ddc2125ecbda7a",
  "title": "DLyte Titanium Acetabular Implant Polishing Case",
  "kind": "reference",
  "company": "DLyte",
  "parameters": {
    "Component": "Titanium acetabular implant",
    "Treated area": "Internal surface only",
    "Electrolyte": "Dry Suspension; L-130 for titanium",
    "Process time": "45 minutes",
    "Batch capacity": "2 pieces",
    "Final roughness": "Below 0.05 µm",
    "Reported material removal": "0.03 g per piece",
    "Media bowl volume": "210 litres",
    "Titanium removal capacity per media charge": "2 kg",
    "Calculated treatment quantity": "More than 65,000 implants per media charge"
  },
  "summary": "A titanium acetabular implant case focused on polishing only the internal surface, reducing roughness below 0.05 µm with low material removal and controlled batch processing.",
  "sections": [
    {
      "heading": "Finishing objective",
      "paragraphs": [
        "The process was defined to polish the internal bearing surface while leaving the exterior treatment boundary controlled.",
        "The documented target was roughness below 50 nm with homogeneous results between batches."
      ]
    },
    {
      "heading": "Validated cycle",
      "paragraphs": [
        "Trial cycles varied time and other process parameters, with profilometer measurements taken in multiple areas. The selected cycle used dry-suspension media.",
        "The recorded production condition processed two implants in 45 minutes and achieved final roughness below 0.05 µm."
      ]
    },
    {
      "heading": "Material-removal accounting",
      "items": [
        "Recorded removal: 0.03 g per implant",
        "Working-bowl volume: 210 L",
        "Titanium removal capacity with L-130 media: 2 kg per charge",
        "The stated capacity implies treatment of more than 65,000 implants before reaching the removal limit."
      ]
    },
    {
      "heading": "Production controls",
      "items": [
        "Mask or fixture untreated zones before processing.",
        "Measure mass and roughness on a defined sampling plan.",
        "Confirm geometry, tolerance and surface cleanliness after each validated media-life interval.",
        "Medical-device release still requires the manufacturer’s approved validation and biocompatibility controls."
      ]
    }
  ],
  "updated_at": "2026-10-13 13:31:12"
}