POLISHX

PRESI specialised polishing equipment selection guide

This guide distinguishes four specialised preparation roles in the PRESI range: gentle vibration super-finishing, portable in-situ mechanical preparation, an ELLOPOL II electropolishing machine identity, and a programmable electrolytic power supply. The units are related by workflow but are not interchangeable.

ManufacturerPRESI
Reference scopeSpecialised polishing-equipment comparison
VIBROTECH 300 reference68100
VIBROTECH diameter305 mm
Electer Evolution reference58100
Electer speed1,500–25,000 rpm
ELLOPOL II reference59004
Power-supply reference59101
Power-supply output0–100 V
Process familiesMechanical; vibration; electrolytic

Equipment-family map

VIBROTECH 300 — ref. 68100305 mm vibration-polishing machine for ultra-soft final preparation, including EBSD and AFM specimens.
Electer Evolution — ref. 58100Portable in-situ mechanical grinding and polishing unit with 1,500–25,000 rpm drive, reversible rotation and reduction gear.
ELLOPOL II — ref. 59004Identified as an electropolishing machine; detailed operating parameters are not stated in the reviewed product page.
Power supply — ref. 59101Programmable 0–100 V unit for electrolytic polishing and etching, with dual current ranges and separate timers.

Choosing by process stage

Use Electer Evolution when a local surface must be mechanically prepared on a large or installed component before etching. Use VIBROTECH 300 when the need is a low-aggression final laboratory polish that preserves inclusions and reduces preparation-induced surface hardening.

For electrolytic work, treat the polishing machine or cell and the electrical supply as separate functional elements unless the configured system documentation states otherwise. Reference 59101 provides controlled power and timing; electrolyte selection and specimen-cell geometry remain method-specific.

Key selection checks

Specimen locationLaboratory specimen, installed component or clamp-held workpiece
Removal modeMechanical abrasion, vibration super-finishing or electrochemical removal
Analytical objectiveRoutine surface preparation, etching, EBSD or AFM
Control requirementMechanical speed and direction versus voltage, current and timed electrochemical cycles
Method validationMaterial-specific abrasives or electrolyte, confirmed settings, cleaning and repeatability records

Safety and upkeep

Mechanical systems require guarding, speed-compatible tooling and isolation before tool changes. Vibration equipment must be stopped before specimens or holders are repositioned.

Electrolytic preparation requires compatible cells, ventilation, electrolyte-specific protective equipment and de-energised connections during setup. Keep mechanical abrasive residue away from electrolyte equipment and document cleaning between methods.

POLISHX technical profile · 2026-10-14