  "title": "Electrolytic polishing power supply 59101",
  "kind": "product",
  "company": "PRESI",
  "category": "Chemical & special polishing",
  "parameters": {
    "Reference": "59101",
    "Input": "220 V single-phase; 50/60 Hz",
    "Output voltage": "0–100 V adjustable",
    "Current ranges": "0–10 A; 0–20 A",
    "Control": "PLC with stored polishing and etching ranges",
    "Timers": "2; polishing and etching",
    "Outputs": "Rear electropolishing unit; front clamp holder"
  },
  "summary": "Reference 59101 is a 220 V single-phase power unit for electrolytic polishing and etching. It provides adjustable 0–100 V DC output, two ammeter ranges, stored PLC process ranges, separate polishing and etching timers, and connections for either an electropolishing unit or a clamp-type work holder.",
  "sections": [
    {
      "heading": "Electrical capabilities",
      "rows": [
        [
          "Input",
          "220 V, single phase, 50 or 60 Hz"
        ],
        [
          "Output",
          "Adjustable from 0 to 100 V with voltage display"
        ],
        [
          "Rectification",
          "Direct current rectified by silicon diodes"
        ],
        [
          "Ammeter ranges",
          "0–10 A and 0–20 A"
        ],
        [
          "Process timing",
          "Separate polishing and etching timers"
        ],
        [
          "Control",
          "PLC-managed programming with memorisation of polishing and etching ranges"
        ]
      ]
    },
    {
      "heading": "Connections and sequencing",
      "paragraphs": [
        "The rear output is intended for connection to an electropolishing unit. A front connector supports work with a special clamp-type holder, allowing the supply to serve different specimen arrangements.",
        "Time-delay functions coordinate startup and electrolyte-flow control. Establish the required electrolyte circulation before applying process output, and use the separate timing functions to distinguish polishing from etching cycles."
      ]
    },
    {
      "heading": "Method setup",
      "paragraphs": [
        "Choose voltage, current range and process time from a validated method for the alloy and electrolyte. Begin with output disabled, connect the workpiece and counter-electrode arrangement, then verify polarity and electrolyte flow before energising.",
        "Monitor the displayed voltage and measured current throughout method development. Store only a confirmed polishing or etching range, and label the saved method so it cannot be confused with a different alloy or electrolyte."
      ]
    },
    {
      "heading": "Electrical and chemical safety",
      "paragraphs": [
        "Electropolishing combines electrical energy with corrosive or otherwise hazardous electrolytes. Use a compatible cell, ventilation, splash protection and personal protective equipment defined by the electrolyte safety data.",
        "Switch the output off before changing specimen connections. Isolate the 220 V input before maintenance, inspect cables and connectors for chemical attack, and do not operate when electrolyte leakage can reach the power unit."
      ]
    }
  ],
  "updated_at": "2026-10-14"
}