  "title": "Rhopoint IQ Flex 20 - Rhopoint Instruments",
  "kind": "product",
  "company": "Rhopoint Instruments",
  "category": "Metrology & quality control",
  "parameters": {
    "sku": "A6000-016.2",
    "category": "Gloss, Haze, DOI &amp; Opacity &gt; Glossmeters for curved surfaces",
    "brand": "Rhopoint Instruments"
  },
  "summary": "Portable 20° appearance instrument for measuring gloss, reflected haze, reflected image quality and distinctness of image on small or curved high-gloss parts. Magnetic spot-size and part adaptors support repeatable positioning on geometries that are difficult for a conventional flat-surface glossmeter.",
  "sections": [
    {
      "heading": "Measured appearance parameters",
      "paragraphs": [
        "The measuring head records 20° gloss together with reflected haze, RSPEC peak reflectance, reflected image quality and distinctness of image. These parameters distinguish loss of image clarity and surface texture effects that a single gloss value may not reveal.",
        "The standard measurement area is 6 × 6.4 mm. Optional magnetic adaptors reduce the spot to 4 mm or 2 mm, and dedicated curved-part or custom adaptors improve positioning on cylindrical and irregular components."
      ]
    },
    {
      "heading": "Range and repeatability",
      "rows": [
        [
          "Gloss range",
          "0–100 GU / 100–2000 GU"
        ],
        [
          "Gloss resolution",
          "0.1 GU"
        ],
        [
          "Gloss repeatability",
          "0.2 GU / 0.2%"
        ],
        [
          "Haze resolution",
          "0.1 HU"
        ],
        [
          "Haze repeatability",
          "±0.5 HU"
        ],
        [
          "RIQ range",
          "0–100"
        ],
        [
          "RIQ repeatability",
          "±0.2"
        ]
      ]
    },
    {
      "heading": "Use and calibration",
      "paragraphs": [
        "Use the same adaptor for calibration and measurement. Recalibrate when changing between the standard and reduced spot sizes; an eight-hour calibration interval is specified for routine work.",
        "For high-gloss plastics and coatings, compare readings with the appropriate 20° reference geometry. Custom or curved-part adaptors are preferable where the specimen cannot sit squarely over the optical head."
      ]
    },
    {
      "heading": "Instrument configuration",
      "rows": [
        [
          "Memory",
          "8 MB / 999 readings"
        ],
        [
          "Power",
          "Rechargeable lithium-ion battery"
        ],
        [
          "Rated operation",
          "6 hours or more than 4,000 readings per charge"
        ],
        [
          "Recharge time",
          "4 hours"
        ],
        [
          "Dimensions",
          "150 × 79 × 34 mm without head"
        ],
        [
          "Weight",
          "550 g"
        ]
      ]
    }
  ],
  "updated_at": "2026-10-14"
}