Page 1 Number Powder Cure Heat Cool Mold Mold Size of Amount Time Time Pressure Diameter Temp Scoops (g) (min) (°C/°F) (min) (psi) 1"&25 mm 1 24 1 4 4400 1.25" & 30 mm 1.5 37 1 4 4000 180/356 1.5" & 40 mm 2 49 1 4 3800 2" &50 mm 3.5 85 1 6 3500 Notes: _ For single mounts. If duplexing, may need to add 1 minute to cure time. = _ Powder amount may vary depending on the size of each sample. All samples must be ultrasonically cleaned and dried to remove any debris and/or grease. Mold Release agent must be applied every time before a mounting cycle (200-10006 PTFE Spray is recommended). Cure cycle times and temperatures may vary for non-Allied presses.