Buehler NeoTerra is an electrolytic metallography system that applies electrical current through a targeted nozzle and uses recirculating, acid-free TerraBlend chemistry to reveal microstructure. The system supports samples up to 152 × 152 × 50 mm and provides a selectable etch area up to 93 × 96 mm.

Technical specifications

Manufacturer / publisherBuehler
CategoryChemical & special polishing
Updated2026-10-13T00:33:15.655Z

Process and workflow

NeoTerra is designed to reduce the number of conventional grinding, polishing and chemical-etching steps. After planar grinding, the sample is placed in the chamber and electrically connected; the operator uses the joystick and laser sight to select the target area, chooses a stored method and starts the cycle.

The nozzle confines the active area, while TerraBlend chemistry recirculates through the machine. Method settings are intended to make repeat preparation more consistent when the same material and target area are processed.

Capacity and installation

  • Maximum sample envelope: 152 × 152 × 50 mm.
  • Maximum selectable etch area: 93 × 96 mm.
  • Machine envelope: 699 mm deep, 889 mm wide and 775 mm high.
  • Electrical input is single phase at 50/60 Hz; maximum connected load is 3000 W.

Consumables and process control

TerraBlend-1 and TerraBlend-2 are the dedicated recirculating consumables. Each refill bottle contains 5 L. The correct blend and stored method must be matched to the sample material; unknown materials should be qualified with a trial before routine processing.

  • TerraBlend-1: 499211; bundle: 499221.
  • TerraBlend-2: 499212; bundle: 499222.
  • Mounted sample holders are available as accessories.

Maintenance and safety

Keep the reservoir, nozzle path, sample connection and working area clean so residues do not affect later specimens. Check consumable condition and electrical contact before each method run; no maintenance interval was specified in the captured technical page.

The acid-free process reduces the need for conventional acid etchants, but the selected TerraBlend still requires its current safety data, local laboratory controls and appropriate personal protection.