THE GLOBAL POLISHING & ABRASIVES DIRECTORY查看数据库覆盖情况 ↗
POLISHXEN
POLISHX/Fuses 1088 Susp Plub — Electronic Components
Technical document

Fuses 1088 Susp Plub — Electronic Components

Struers methods: Fuses 1088 Susp Plub. Source classification: Electronic Components.

Publication details

Manufacturer / publisherStruers
EditionListed by source
Material familyElectronic Components
CategoryLapping, honing & superfinishing
Updated2026-10-17
Struers published content
Source: https://publications.struers.com/brochures/english/application-notes/methods/fuses-1088-susp-plubelectronic-components/ · Original publication: . Manufacturer claims and terminology are reproduced as source material.Explore Preparation methods ↗

Struers methods: Fuses 1088 Susp Plub. Source classification: Electronic Components.

Original preparation method, page 1
Original method · Page 1
Searchable text transcription

The page images and original PDF preserve the table layout and notation.

Grinding
Steps
PG
FG
FG 2
FG 3
Surface
Type
SiC Foil #320
SiC Foil #500
SiC Foil #1200
SiC Foil #4000
Disc speed (rpm)
300
300
300
300
Abrasive
Type 
Predosing/dosing
Lubricant
Type 
Water
Water
Water
Water
Predosing/dosing
Force (N)
15
15
15
15
Holder direction
>>
>>
>>
>>
Holder speed (rpm)
150
150
150
150
Time (min)
1
1
1
2
Fuses
1088
Electronic Components 
Susp. P/Lub.
Polishing
Steps
DP
OP
Surface
Type
MD-Mol
MD-Chem
Disc speed (rpm)
150
150
Abrasive
Type 
DP-Susp. P 3 µm
OP-U, 0.04 µm
Predosing/dosing
3/6
1/10
Lubricant
Type 
DP-Blue
Predosing/dosing
3/12
Force (N)
20
10
Holder direction
>>
><
Holder speed (rpm)
150
150
Time (min)
6
1
The method is developed to fit a standard equipment 
configuration with 30 mm specimen on 300mm disc. 
Dosing levels are valid for following equipment: 
Tegramin, AbraPol and LaboPol
For advice on how to adjust the method to fit your 
equipment, please take a look at Preparation Parameters.
The preparation time for the plane grinding (PG) step 
depends very much on the condition of the samples and 
if they are prepared as single specimens or clamped in a 
specimen holder. Please check the samples after the first 
step and repeat it if necessary.
Method comments
Take a look at Preparation Parameters 

Files & downloads

fuses-1088-susp-plub.pdf ↓

Related products & technical information