Phenolic Mounting Powder Cure Schedule
Full searchable text · 1 pages
Refer to the original PDF for tables, figures and page layout.
Automatically recognized text. Verify numerical values against the original PDF.
Page 1 Cure Schedule — Phenolic Mold Size Number Powder Cure Time Temp Cool Time] Mold Pressure Diameter of Scoops | Amount (g) | (min:sec) (°C/°F) (min) (psi) 1"&25 mm 1 11-13 4 4400 1.25" & 30 mm 2 23-25 4 4000 200/392 1.5" &40 mm 2.5 28-31 3800 2" & 50 mm 4 45-50 6:30 6:30 3500 Notes: = For single mounts. If duplexing, 1 minute may need to be added to the cure time. =a = Powder amount may vary depending upon the size of each sample. All samples must be ultrasonically cleaned and dried to remove any debris and/or grease. = A mold release agent must be applied every time before a mounting cycle; Allied’s mold release for hot mounting (#200-10004) is recommended. Cure cycle times and temperatures may vary for non-Allied presses.