Electrolytic polishing power supply 59101
Reference 59101 is a 220 V single-phase power unit for electrolytic polishing and etching. It provides adjustable 0–100 V DC output, two ammeter ranges, stored PLC process ranges, separate polishing and etching timers, and connections for either an electropolishing unit or a clamp-type work holder.
Technical specifications
| Manufacturer / publisher | PRESI |
|---|---|
| Reference | 59101 |
| Input | 220 V single-phase; 50/60 Hz |
| Output voltage | 0–100 V adjustable |
| Current ranges | 0–10 A; 0–20 A |
| Control | PLC with stored polishing and etching ranges |
| Timers | 2; polishing and etching |
| Outputs | Rear electropolishing unit; front clamp holder |
| Category | Chemical & special polishing |
| Updated | 2026-10-14 |
Electrical capabilities
| Input | 220 V, single phase, 50 or 60 Hz |
|---|---|
| Output | Adjustable from 0 to 100 V with voltage display |
| Rectification | Direct current rectified by silicon diodes |
| Ammeter ranges | 0–10 A and 0–20 A |
| Process timing | Separate polishing and etching timers |
| Control | PLC-managed programming with memorisation of polishing and etching ranges |
Connections and sequencing
The rear output is intended for connection to an electropolishing unit. A front connector supports work with a special clamp-type holder, allowing the supply to serve different specimen arrangements.
Time-delay functions coordinate startup and electrolyte-flow control. Establish the required electrolyte circulation before applying process output, and use the separate timing functions to distinguish polishing from etching cycles.
Method setup
Choose voltage, current range and process time from a validated method for the alloy and electrolyte. Begin with output disabled, connect the workpiece and counter-electrode arrangement, then verify polarity and electrolyte flow before energising.
Monitor the displayed voltage and measured current throughout method development. Store only a confirmed polishing or etching range, and label the saved method so it cannot be confused with a different alloy or electrolyte.
Electrical and chemical safety
Electropolishing combines electrical energy with corrosive or otherwise hazardous electrolytes. Use a compatible cell, ventilation, splash protection and personal protective equipment defined by the electrolyte safety data.
Switch the output off before changing specimen connections. Isolate the 220 V input before maintenance, inspect cables and connectors for chemical attack, and do not operate when electrolyte leakage can reach the power unit.